| Display |
10.4” LCD Display |
| Mode: |
a pneumatically powered and electrically controlled system |
| Trolley: |
Fitted with 4 nos anti-static rubber castors; two of which
are lockable for braking and easy maneuverability with foot-operated brake provisions
|
| Technical Specifications: |
| Gas requirement: |
Medical oxygen and nitrous oxide with a pressure ranging from 280~600kpa |
| safety valve |
<12.5 kPa |
| Respiratory Rate |
1 ~ 99 bpm |
| Oxygen Concentration |
21% ~ 100% |
| Maximum Minute ventilation |
≥18L / min |
| modes of ventilation |
A/C, IPPV, SIPPV, IMV, SIMV, PEET. MANUAL, SIGH, SPONT |
| Ventilator compliance |
≤30 Ml/kPa |
| PEEP: |
0 ~ 1.0 kPa |
| Ptr: |
-10 ~ 10 hPa |
| IMV frequency: |
1~20 bpm |
| I/E Ratio: |
4:1~1:4, Has the inverse ratio ventilation |
| Tidal Volume |
Setting Range: 50~1500ML
Display Range: 0~2000ml ; resolution ratio: 10ml
|
| Inspiratory Plateau: |
0~1s |
| Maximum Safety Pressure |
≤ 6.0kPa |
| Sigh: |
one deep breath per 80 controlled respirations,
the inspiration time is 1.5 times of the setting point
|
| Pressure limit range: |
1.0~ 6.0kPa |
| Airway pressure alarm: Audible and visual and
with yellow and red color indicating
|
Lower: 0.1 kPa ~ 6.0 kPa; Upper: 0 ~ 5.0 kPa |
| ±0.2 kPa |
| Ventilation Volume alarm: |
≥ 18L/ MIN |
| Power Supply Alarm |
Ac/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s |
|
| Operating conditions |
| Ambient temperature: |
10 ~ 40oC |
| Relative humidity: |
no higher than 80 % |
| Atmospheric pressure: |
860 hPa ~ 1060 hPa |
| Power requirement: |
100-120 Vac, 50/60 Hz; |
| Attention: the AC power supply used for the anesthesia
the machine must be well-grounded.
|
|
| Storage |
| Ambient temperature: |
-15oC ~ +50oC |
| Relative humidity: |
no higher than 95 % |
| Atmospheric pressure: |
86 kPa ~ 106 kPa. |
|
It should be stored in a room without corrosive gas and well ventilated |
| Package |
| packaging box |
comply with the requirement of GB/T 15464 |
|
Between the packaging box and product, soft material
with appropriate thickness provided to prevent loosening
and mutual friction during transportation
|
|
Moisture protection and rain protection to ensure that
product is protected from natural damage.
|
| Safety & Alarm |
| Gas source pressure |
Immediate alarm; “high level alarm” indicator flickers raising an audible alarm. |
| Airway Pressure Alarm |
The airway pressure exceeds 15hpa. The maximum delay is no longer than 17s. The “high-level alarm” indicator flickers raising an audible alarm. |
| Tidal Volume Alarm |
Lower than 50ml, The alarm will be raised after one breathing cycle. |
| Ventilation Volume Alarm |
Lower: 0~12L/Min; Up: 18L/ Min |
| Battery Voltage Alarm |
It alarms when the battery voltage is lower than 10.5V. |
| Power On Self Test |
Self-test before working, assure it’s working normally. |